Wide Dual Mini-VerticalsMini-Vertical Slab Gel /Blotting SystemDCX-800C, cooledDCX-800INSTRUCTION MANUAL
10Wide Dual Mini-Vertical Instructions 9-30-133.2 Running the Gel1. Place stirring bar into bottom of reservoir in stirring corral (as shown in gure
11www.cbsscientic.commatching colors3.3 Removing the Gel1. Turn the power supply off and disconnect the leads from the power supply. Remove the safe
12Wide Dual Mini-Vertical Instructions 9-30-134.1 Alternate Protocol: Using Gel Wrap™ Gasket Casting Method (Not using Pre-cast gels)1. Place all c
13www.cbsscientic.com4.1 Alternate Protocol: Using Gel Wrap™ Gasket Casting Method (Not using Pre-cast gels)1. Place all components in an authoriz
14Wide Dual Mini-Vertical Instructions 9-30-13SECTION 4 Alternate Protocol for Slab Gel Electrophoresis. (Not using Pre-cast gels)4.2 Preparing the
15www.cbsscientic.comSECTION 5 Instructions for Western Blotting5.1 Preparing the Unit for Blotting1. Remove safety cover from the assembled unit by
16Wide Dual Mini-Vertical Instructions 9-30-134. Insert blotting casettes into core making sure that red side faces outward. See diagram 5.5. Close do
17www.cbsscientic.com4. Insert blotting casettes into core making sure that red side faces outward. See diagram 5.5. Close doors and re-latch by pres
18Wide Dual Mini-Vertical Instructions 9-30-136.2 Electro-Blotting6.3 References1. Hames, B.D. (ed.) (1998). Gel Electrophoresis of Proteins. A Pra
19www.cbsscientic.com6.4 Recommended ClearPAGE™ Buffer FormulationsAs an alternative to the ClearPAGE buffers available for purchase, these formul
2Wide Dual Mini-Vertical Instructions 9-30-13TABLE OF CONTENTS Important User Information . .
20Wide Dual Mini-Vertical Instructions 9-30-136.4 Recommended ClearPAGE™ Buffer Formulations, con’t.LDS Sample Buffer, 4x, (Cat. # FB31010) *pH
21www.cbsscientic.comSECTION 7Maintenance of Equipment7.1 Care and Handling The plastic components of the Wide Dual Mini-Vertical Electrophoresis Sy
22Wide Dual Mini-Vertical Instructions 9-30-13SECTION 8Ordering InformationCOMBS: See page 6.Wide Dual Mini-Vertical Ordering Information CAT. # DES
23www.cbsscientic.comClearPAGE Precast Gel Selection Chart & Reagent Ordering information CAT. # DESCRIPTIONClearPAGE™ BUFFERS, STAI
CONTACT INFORMATIONOnline Ordering www.cbsscientic.comTelephone Local and International: 858-755-4959 Toll Free: 800-243-4959Sales E-mail Address
3www.cbsscientic.comIMPORTANT USER INFORMATIONThis Instruction Manual will explain how to use this product safely and effectively. Please read and c
4Wide Dual Mini-Vertical Instructions 9-30-13FRANÇAIS INFORMATION IMPORTANTE À L’USAGE DES UTILISATEURSLe présent manuel d’utilisation explique la
5www.cbsscientic.comSECTION 1General Information 1.1 Introduction and FeaturesC.B.S. Scientic offers the Wide Dual Mini-Verticals are for performin
6Wide Dual Mini-Vertical Instructions 9-30-131.3 Comb Specications for Hand Cast Gels1.4 SafetyPower to the Wide Dual Mini-Vertical Electrophores
7www.cbsscientic.comSafety CoverCoreTubing adaptors(DCX-800C only)SECTION 2 Description of Parts2.1 Unpacking and Components Please verify that y
8Wide Dual Mini-Vertical Instructions 9-30-13SECTION 3 Instructions for Slab Gel Electrophoresis Using Pre-Cast Gels3.1 Preparing the Electrophoresi
9www.cbsscientic.com45. Slide pre-cast gel cassette or plate set(s) into the core assembly with the notched plate facing in towards the upper buffer